Using Transformers for Electrical Isolation in Medical Devices

Toroidal transformers are ideal for electrical isolation in medical equipment because they are compact, can be completely encapsulated when necessary, and have low stray fields, making them less likely to cause electromagnetic disturbances.

Isolation means physically and electrically separating two parts of a circuit that may interact. The isolation is achieved by using electromagnetic field coupling between the two circuits.

For an isolation method to be safe enough to use in medical equipment, it needs to have two things: high-integrity isolation components and a safe insulator barrier. For example, an insulator can be a piece of plastic, a keep-out space in a PWB, or an air gap.

The three most commonly used isolation methods are optocouplers (light), transformers (magnetic flux), and capacitive couplers (electric field).

Isolation provides several safety advantages in medical equipment:

  • it breaks ground loops
  • it improves common-mode voltage rejection
  • it permits the two parts of the circuit to be at different voltage levels, which means one side can be safe while the other side is at hazardous voltage levels.

For the purposes of this discussion, we will focus on using transformers as a method of electrically isolating medical equipment.

Transformer Overview

A transformateur is an electrical device that transfers energy between two or more circuits through electromagnetic induction.

Commonly, transformers are used to increase or decrease the voltages of alternating current in electric power applications. This is accomplished by passing a varying current through the primary winding to generate a magnetic flux in the transformer’s core.

This flux then induces a voltage in the transformer’s secondary winding. Depending on the ratio of the primary windings to the secondary windings, the transformer’s output voltage can be increased or decreased.

For most transformers designed for use within homes or offices, one of two styles of transformers is used: EI laminate core or toroidal core.

E-I Transformer Construction

In an E-I structure, the matching “E” and “I” components are stamped from sheets of thin grain-oriented electrical steel that are later stacked to create the core. The primary and secondary windings are wound on bobbins. Multiple bobbins are placed on spindles and spun in order to apply the windings.

This method of using bobbins allows for automation that reduces the manufacturing times and also provides insulation between the windings and the core. The EI core laminations are stacked inside the bobbins to complete the transformer.

Toroidal Transformer Construction

A toroidal core is made from a continuous strip of silicon steel, which is wound like a tight clock spring. The ends are tacked into place with small spot welds, to prevent the coiled steel from unwinding. The core is insulated with an epoxy coating or a set of caps or multiple wraps of insulating film (such as Mylar or Nomex).

The transformer’s windings are applied directly onto the core itself. Additional insulation is required to isolate the windings.

Since the windings must be wound through the center hole of the core and the core is one piece, bobbins can’t be used on toroidal transformers. This makes manufacturing toroidal transformers more labor-intensive.

If toroidal transformers are more labor-intensive and do not lend themselves to automation, why are they used? The answer lies in performance.

The continuous strip of steel used in the core allows the transformer to be smaller, lighter, more efficient, and quieter than an EI laminate. These qualities are highly desirable in medical electrical equipment (and many other applications) and justify the additional expense.

Conclusion

Toroidal transformers are ideal for electrical isolation in medical equipment because they are compact and can be completely encapsulated when necessary. They also have low stray-fields and are therefore less likely to cause radiated electromagnetic disturbances.

In our next article, we will dive deeper into toroidal medical isolation transformers and their design.


Learn more in our white paper — Improving Medical Isolation Transformers with Segmented Core Cap Technology